SGI Ethernet Card Dokumentacja Strona 138

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 154
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 137
Calibration SG Series Programming
7-16 M550129-03 Rev G
6. Verify that the unit is set to 100% ± 0.25% of full-scale output voltage. If
necessary adjust R39 (on the isolated interface adapter) for 100% output
voltage.
7. Set the programming source to 0.5V ±1mV.
8. Verify that the unit is set to 10% ± 0.25% of full-scale output voltage. If
necessary adjust R35 (on the isolated interface adapter) for 10% output voltage.
9. Repeat the steps above as required to obtain the required accuracy.
10. Remove all connections to the remote analog control connector, J1 (Figure 7-6).
Place the jumper between pins 5 and 6. This reverts the SGI series unit back to
local, front panel control.
Figure 7-6. Analog Control Connector (J1)
Przeglądanie stron 137
1 2 ... 133 134 135 136 137 138 139 140 141 142 143 ... 153 154

Komentarze do niniejszej Instrukcji

Brak uwag